GK-1000 filter comprehensive performance tester
Product model:
Product Features
Product Details
Introduction to 1.
Standard reference
(1) YY0469-2011 "medical surgical masks"
(2) GB2626-2006 "respiratory protective equipment self-priming filter type anti-particulate matter respirator"
(3) The European StandardEN1822-3:2009 Single Filter Material Test
(4) GB19083-2010 medical protective mask
(5) GB/T 32610-2016 (Technical specification for daily protective masks)
(B) GB/T 19082-2009 Medical Disposable Protective Clothing
The test bench uses polydisperse experimental aerosols (DEHSorNaCl2) Quantitative dust generation, and under the action of clean air flow (negative pressure type), the regular performance test shall be carried out through aerosol particle sampling and pressure difference measuring points upstream and downstream of the tested filter material.The test bench refers to the advanced design concept of similar international equipment (such as the United States.TSI company 8130 models), but also suitable for the domestic related industry testing status quo. Independent research and development of the program software is stable and reliable, easy to operate. Simple running interface, running parameters at a glance. Different dust sources (DEHSandNaCl)Switch freely, suitable for a variety of air filter materials (melt blown, glass fiber,PP, film and other filter materials), medical masks and some labor supplies testing.
2. basic configuration
1.DEHS aerosol generator
2.NaCl aerosol generator.0
3. Laser particle counter
4. Differential pressure transmitter
5. Flow transmitter
6. Spray concentration adjustment
7. High pressure fan
8. Spray air pump
9. Electrostatic neutralization device
10. Electric heating device
11. Standard fixture (100 cm2)
12. Programmable controller
13. Frequency converter
14. Industrial control touch screen
15. Built-in micro printer
16. Miniature oil-free air compressor(Optional)
3. technical parameters
1. Applicable filter material:(1) Glass fiber filter material;(2) PP filter material; (3) film-coated filter material
2. Detect particle size:0.3、0.5、1.0、2.0,5.0、10. 0 μm (6th gear)
3. Flow range:20~100 L/min
4. Detection resistance range:0~1000 Pa
5. Dust source: (1) Aerosol (PA0-4);(2) Sodium salt (2% NaCl solution)
6. Standard test area: 100 cm2
7. Power supply:220V/50 Hz
8. System power consumption: <1500In
9. Overall dimension:1000 mm (width)×680 mm (deep)×1520 mm (height)